Related Experiment Video
Updated: Jul 17, 2026

A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Chronotropic incompetence in patients with an implantable cardioverter defibrillator: prevalence and predicting
Christoph Melzer1, Marco Böhm, Hans Joachim Bondke
1Med. Klinik mit Schwerpunkt Kardiologie, Angiologie, Pneumologie, Charité University Medical Center, Berlin, Germany. christoph.melzer@charite.de
Insights
A significant percentage of implantable cardioverter-defibrillator (ICD) patients exhibit chronotropic incompetence (CI), leading to reduced exercise capacity. This highlights the need to explore rate-adaptive pacing modes for improved physical stress tolerance in ICD recipients.
Area of Science:
- Cardiology
- Electrophysiology
- Exercise Physiology
Background:
- Chronotropic incompetence (CI) is not well-studied in implantable cardioverter-defibrillator (ICD) patients.
- CI may impact ICD device programming and patient outcomes.
- Understanding CI prevalence and predictors in ICD patients is crucial.
Purpose of the Study:
- To investigate the prevalence and predictors of chronotropic incompetence in ICD patients.
- To assess the impact of CI on exercise capacity in this population.
- To inform potential programming strategies for ICDs.
Main Methods:
- 123 ICD patients were categorized into groups based on device type and rhythm.
- Treadmill exercise testing measured heart rate response, VO2max, and anaerobic threshold.
- Clinical data including medications, heart disease, and LV ejection fraction were collected.
Main Results:
- 38% of ICD patients (47/123) demonstrated chronotropic incompetence.
- Predictors of CI included coronary disease, prior cardiac surgery, beta-blocker use, and amiodarone.
- CI was associated with reduced exercise capacity and VO2max.
Conclusions:
- A substantial proportion of ICD patients experience CI, resulting in diminished physical stress tolerance.
- Optimized programming of rate-adaptive modes, considering AV interval and pacing algorithms, is warranted.
- Further research into managing CI in ICD patients may improve quality of life.
Abstract:
Chronotropic incompetence (CI), which has not been systematically examined in the ICD patient population, may have implications for device programming. A total of 123 ICD patients were classified into three groups: single-chamber ICD with sinus rhythm, dual-chamber ICD with sinus rhythm, and single-chamber ICD with permanent atrial fibrillation. Heart rate response, maximum oxygen uptake, and oxygen uptake at the anaerobic threshold were measured during treadmill exercise testing. In addition, clinical variables such as antiarrhythmic drug therapy, underlying heart disease, and left-ventricular (LV) ejection fraction were recorded. Of the patients studied, 38% were chronotropically incompetent (47/123). Significant predictors of CI were as follows: presence of a coronary disease (P = 0.036), prior cardiac surgery (P = 0.037), chronic drug therapy with beta-blockers (P = 0.032), administration of amiodarone (P = 0.025), and a combination of these two forms of treatment (P = 0.01). Spiroergometry revealed reduced exercise capacity (P = 0.041) and lessened VO2max (P = 0.034) among chronotropically incompetent patients. A large percentage of ICD patients demonstrates CI with subsequently reduced physical stress tolerance. In light of the DAVID study, we believe that a closer examination of rate-adaptive modes for ICD patients is warranted under enhanced conditions: (1) optimized AV interval programming; (2) utilization of new algorithms to reduce ventricular pacing in combination with rate-adaptive atrial pacing, with the goal of addressing CI while minimizing ventricular pacing; and (3) an optimized upper heart-rate limit.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
05:16Cutoff Value of Phase Angle by Bioelectrical Impedance Analysis at Admission as a Prognostic Factor in Patients with Acute Heart Failure
Published on: June 10, 2025
Related Concept Videos
Decreased pulse rate
There are specific risk factors that can elevate the likelihood of developing bradycardia. Advanced age is a significant factor, with bradycardia...
Mechanism of Cardiac Arrhythmias
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Cardiomyopathy II: Dilated Cardiomyopathy
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias