Related Experiment Video
Updated: Aug 15, 2026

Characterization of the Isolated, Ventilated, and Instrumented Mouse Lung Perfused with Pulsatile Flow
Published on: April 29, 2011
Comparison of two flow generators with a noninvasive ventilator to deliver continuous positive airway pressure: a
Carolina Fu1, Pedro Caruso, Jeanette Janaina Jaber Lucatto
1Physiotherapy, Phonology and Occupational Therapy Department, Medical School, University of São Paulo, São Paulo, Brazil.
Objective:
To compare the performance of two continuous flow generators with a ventilator designed for noninvasive positive pressure ventilation (NPPV) to deliver continuous positive airway pressure (CPAP). The performance of flow generators using different oxygen pressure supplies was also compared.
Design And Setting:
Experimental study using a mechanical lung model in a university research laboratory.
Measurements:
Two flow generators supplied at 100, 200, and 300 kPa and an NPPV ventilator were compared at CPAP of 5, 10, and 15 cmH2O in: (a) area under the adjusted CPAP level during inspiration, (b) capacity to attain the preset CPAP, and (c) tidal volume.
Results:
The NPPV ventilator attained the preset CPAP better than flow generators, but its area under adjusted CPAP was similar to or higher than that of flow generators when these were adjusted to their better pressure supply. Both flow generators had better performance with an output flow around 100 l/min, which was achieved at 100 kPa with one flow generator and 300 with the other. Flow generators and the NPPV ventilator generated similar tidal volumes.
Conclusions:
Flow generators performance showed large variations among different devices and oxygen pressure supplies. Adjusted to their better pressure supply, flow generators had a similar or better capacity to maintain the CPAP level, but the NPPV ventilator was more reliable to attain the preset CPAP. Flow generators could be an alternative to provide CPAP in low-income areas, usually with scarce medical equipment availability.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation I: Indication and Settings
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Pulmonary Function Tests
Pulmonary Function Tests are crucial diagnostic tools for assessing respiratory function, particularly in patients with chronic respiratory disorders. They comprehensively evaluate lung volumes, ventilatory function, breathing mechanics, diffusion, and gas exchange. These tests help diagnose pulmonary diseases and play a significant role in monitoring disease progression, evaluating disability, and assessing response to therapy.
PFTs involve using a spirometer, a...