An ex-vivo experimental study on optimization of bipolar radiofrequency liver ablation using perfusion-cooled
1Radiology and Institute of Radiation Medicine, Seoul National University College of Medicine, Republic of Korea. leejm@radcom.snu.ac.kr
Purpose:
To determine optimal parameters for bipolar radiofrequency ablation (RFA) using perfusion-cooled electrodes to create a large ablation volume in ex vivo bovine liver.
Material And Methods:
Three sets of RF experiments were performed using a 200-Watt generator and two 15-gauge perfusion-cooled or internally cooled electrodes in ex vivo bovine livers. In the first set of experiments, to find the ideal inter-electrode distance for creating large coagulation necrosis, 30 ablation lesions were created by bipolar RFAs at inter-electrode spacings of 3 cm, 4 cm, and 5 cm. In the second set of experiments, to explore the ideal duration of RF application, bipolar RFAs were performed for 10 min and 20 min. In the first and second experiments, 10 lesions were made for each condition with infusion of 6% hypertonic saline (HS) at 2 ml/min. In the third set of experiments, 10 ablation lesions were created by bipolar RFAs using internally cooled electrodes without HS infusion. The mean volume of those ablation lesions was then compared to that of the lesions created by bipolar RFA using perfusion-cooled electrodes in the second experiments. Tissue impedance, dimension, and shape of the ablated areas were compared in each condition.
Results:
In the first set of experiments, bipolar RFA created a homogeneous oval or spherical-shaped ablation area between the electrodes at 3-5 cm spacing, but showed a more spherical-shaped lesion at 3 cm inter-electrode spacing than at 4 cm and 5 cm spacing. In the second set of experiments, RF energy delivered for 20 min created a larger dimension of coagulation necrosis than energy delivered for 10 min: 107.6 +/- 34 cm3 versus 59.5 +/- 27 cm3 (P<0.05). In addition, the mean volume of ablation regions obtained with bipolar RFA using the internally cooled electrode was 47.5+/- 17 cm3, which was significantly less than that with bipolar RFA using perfusion-cooled electrodes (P <0.05).
Conclusion:
Bipolar RFA using perfusion-cooled electrodes achieves homogeneous areas of coagulation necrosis between two electrodes, preferably at 3 or 4 cm inter-electrode distance for 20 min, and is better in creating large coagulation necrosis than bipolar RFA using internally cooled electrodes.