Medical nebulizer performance: effects of cascade impactor temperature
Yue Zhou1, Amitkumar Ahuja, Clinton M Irvin
1Lovelace Respiratory Research Institute, 2425 Ridgecrest Drive SE, Albuquerque NM 87108, USA. yzhou@lrri.org
Background:
During operation of a jet nebulizer, the temperature of the nebulizer outlet could decrease by more than 10 degrees C, depending on the nebulizer type and operating conditions, such as driving flow rate and fill volume. The droplet size distribution generated from the nebulizer can be measured by a cascade impactor. However, when the cascade impactor is operated at ambient room temperature, the droplets could evaporate because of the temperature difference between the nebulizer outlet and the body of the impactor.
Methods:
An 8-stage cascade impactor was used to measure the particle size distribution from 4 different types of jet nebulizer (LC Plus, Side-Stream, VixOne, and Micromist) in 2 temperature conditions: ambient (22 degrees C) and low (10 degrees C). Two different formulations, albuterol (aqueous solution) and budesonide (suspension), were used.
Results:
There was a significantly larger (p < 0.05) mass median aerodynamic diameter and smaller respirable fraction for each nebulizer with the impactor at low temperature than with the impactor at ambient temperature. The mass median aerodynamic diameter of the nebulizers with the impactor operating at low temperature appeared 15-130% larger than with the impactor operating at ambient temperature, for both formulations. The respirable fraction also changed from 10% when the impactor was operated at low temperature to 65% when the impactor was operated at ambient temperature.
Conclusion:
The results provide important information for the use of a cascade impactor to measure the particle-size distribution of nebulizer aerosols.
Related Concept Videos
Inhaled Medications
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation (NIPPV)
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...