Determination of several psychiatric drugs in whole blood using capillary gas-liquid chromatography with nitrogen
Carolina Sánchez de la Torre1, María A Martínez, Elena Almarza
1National Institute of Toxicology and Forensic Sciences, Ministry of Justice, C/Luis Cabrera 9, 28002 Madrid, Spain.
Abstract:
A simple and reliable gas chromatographic method with nitrogen-phosphorus detection without derivatization was developed for the detection of several psychiatric drugs in whole blood as part of systematic toxicological analyses (STA). Drugs included mirtazapine, chlorpromazine, methotrimeprazine (levomepromazine), clothiapine, olanzapine, clozapine, haloperidol, and thioridazine. All drugs were studied at concentrations of 100-2,000 microg/L, except haloperidol that was studied at concentrations of 400-8,000 microg/L. In order to select the best blood purification procedure and therefore increase the signal to noise ratio we have compared two solid-phase extraction (SPE) columns, Chem Elut and Bond Elut Certify, for their recovery, precision, sensitivity and matrix purification efficiency. Recoveries for these drugs using Chem Elut columns at 500 and 2,000 microg/L (2,000 and 8,000 microg/L for haloperidol) were in the range 21-65%, with intra-assay and inter-assay precisions of less than 17% and 19%, respectively. Limits of detection (LODs) and limits of quantitation (LOQs) for mirtazapine, chlorpromazine, methotrimeprazine, clothiapine, olanzapine, clozapine, and thioridazine ranged from 62 to 161 microg/L and from 205 to 531 microg/L, respectively. LOD and LOQ for haloperidol were 442 and 1,458 microg/L, respectively. Recoveries of these compounds using Bond Elut Certify columns at 500 and 2,000 microg/L (2,000 and 8,000 microg/L for haloperidol) were in the range 44-97%, with intra-assay and inter-assay precisions of less than 7% and 14%, respectively. LODs and LOQs for mirtazapine, chlorpromazine, methotrimeprazine, clothiapine, olanzapine, clozapine, and thioridazine ranged from 37 to 66 microg/L and from 122 to 218 microg/L, respectively. LOD and LOQ for haloperidol were 156 and 515 microg/L, respectively. Linearity was observed in the studied range for all compounds with r(2) values of >0.999. The use of the mixed-mode bonded-silica Bond Elut Certify columns showed advantages comparing with Chem Elut columns for the screening of these psychotropic agents such as higher recoveries, cleaner extracts, better sensitivity, better precision and less solvent consumption and subsequent disposal.
More Related Videos
13:35A Convenient Method for Extraction and Analysis with High-Pressure Liquid Chromatography of Catecholamine Neurotransmitters and Their Metabolites
Published on: March 1, 2018
05:31Gas Chromatography-Mass Spectrometry Paired with Total Vaporization Solid-Phase Microextraction as a Forensic Tool
Published on: May 25, 2021
Related Concept Videos
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
High-Performance Liquid Chromatography: Types of Detectors
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...