Related Experiment Video
Updated: Aug 15, 2026

06:58
Endotracheal Intubation in Mice via Direct Laryngoscopy Using an Otoscope
Published on: April 5, 2014
[Electric short-circuit incident observed with "Upsher" laryngoscopes]
1Clinique de l'Orangerie, 29, allée de la Robertsau, 67000 Strasbourg, France. laurent.tritsch@sdis67.com
Annales Francaises D'Anesthesie Et De Reanimation
|October 18, 2005
Abstract:
We observed an electrical short-circuit between a fasten screw of the printed circuit and the handle of an Upsher universal laryngoscope (serial number UQ1). The isolating Silicone layer was broken above the screw. This isolation defect was found all over our Upsher laryngoscopes of the UQ1 series. No doubt that if accumulators were used instead of batteries, emitted heat would be in largest amount and perhaps dangerous.