Quantitative concentration measurements of creatinine dissolved in water and urine using Raman spectroscopy and a
1University of Rochester, The Institute of Optics, Department of Physics and Astronomy, Rochester, New York 14627, USA.
Abstract:
We describe the use of a Teflon-AF liquid core optical fiber (LCOF) geometry to enhance the collection of Raman scattering from the biochemical creatinine, dissolved in water and in urine. At short integration times, where shot noise is most troublesome, the enhanced signal leads to greater accuracy in estimating the creatinine concentration from the spectrum. At longer integration times, instabilities in the LCOF geometry manifest themselves, and the predictions are the same as or worse than those from standard cuvette-based spectral measurements. Photobleaching of fluorescence from urine is more extensive and more stable in the LCOF as well. Starting from the measured enhancement of a major creatinine Raman band, we calculate the expected ratio of prediction errors obtained using the two geometries, and it agrees closely with the observed ratio. These results indicate that Raman spectroscopy with these Teflon-AF LCOFs is stable enough for quantitative concentration predictions, accurate to a few percent of the concentration range spanned.
More Related Videos
09:16Highly Sensitive Measurement of Glomerular Permeability in Mice with Fluorescein Isothiocyanate-polysucrose 70
Published on: August 9, 2019
09:32Resolving Water, Proteins, and Lipids from In Vivo Confocal Raman Spectra of Stratum Corneum through a Chemometric Approach
Published on: September 26, 2019
Related Concept Videos
Raman Spectroscopy: Overview
However, a small fraction of the scattered light exhibits a frequency shift due to the exchange of energy between the incident photons and the...
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...
Drug Dosing in Renal Diseases: Measurement of Glomerular Filtration Rate
Drug Dosing in Renal Diseases: Measurement of Serum Creatinine Concentration and Clearance
Drug Dosing in Renal Diseases: Estimation of Glomerular Filtration Rate Based on Serum Creatinine Concentration
Serum Studies: Renal Function Tests