Related Experiment Video
Updated: Aug 15, 2026

Intravascular Ultrasound Image-Based Finite Element Modeling Approach for Quantifying In Vivo Mechanical Properties of Human Coronary Artery
Published on: December 6, 2024
Effect of varying slice thickness on coronary calcium scoring with multislice computed tomography in vitro and in
Georg Mühlenbruch1, Christoph Thomas, Joachim E Wildberger
1Department of Diagnostic Radiology, University Hospital (RWTH), Aachen, Aachen, Germany. gmuhlen@rad.rwth-aachen.de
Objectives:
To compare coronary calcium scoring results (calcium volume, calcium mass, Agatston score, and number of lesions) of different slice thicknesses using a 16-slice CT (MSCT) scanner.
Materials And Methods:
A nonmoving anthropomorphic thorax phantom with calcium cylinders of different sizes and densities was scanned 30 times with repositioning applying a standardized retrospectively ECG-gated MSCT (SOMATOM Sensation 16; Siemens, Forchheim, Germany) scan protocol: collimation 12 x 0.75 mm, tube voltage 120 kV, effective tube current time-product 133 mAs(eff). Fifty patients (29 male; age 57.2 +/- 8.4 years) underwent a nonenhanced scan applying the same scan protocol. Two image sets (effective slice thicknesses 3 mm and 1 mm) were reconstructed at 60% of the RR interval. Image noise was measured in both studies. Calcium volume, calcium mass and Agatston score were calculated using a commercially available software tool.
Results:
Due to increased image noise in thinner slices, calcium scoring in all scans was performed applying a scoring threshold of 350 HU. In the phantom study, 1-mm slices showed significantly higher scoring results in respect to calcium volume (+8.2%), calcium mass (+12.5%), and Agatston score (+5.3%) (all P < 0.0001). In the patient study, 27 patients had coronary calcifications in 3-mm slices, and 31 patients had coronary calcifications in 1-mm slices. Thinner slices showed significantly higher scoring results in respect to volume (+47.1%), mass (+47.2%), and Agatston score (+29.7%) (all P < 0.0001).
Conclusions:
When comparing 3-mm and 1-mm slices in coronary calcium scoring in MSCT, thinner slices lead to significantly increased scoring results.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...