Stroke volume during orthostatic challenge: comparison of two non-invasive methods
Donald F Doerr1, Duane A Ratliff, Joseph Sithole
1Technology Implementation Branch, NASA, Kennedy Space Center, FL, USA.
Background:
Real time non-invasive determination of stroke volume (SV) is important to astronaut orthostatic testing. We compared simultaneous estimates of SV calculated from peripheral pulse waveforms with a more conventional non-invasive technique.
Methods:
Ten men and nine women completed 12-min protocols. The relative change (% delta) in beat-to-beat SV was estimated non-invasively from changes in pulse waveforms measured by application of infrared finger photoplethysmography (IFP) and thoracic impedance cardiography (TIC). The % deltaSV values were calculated from continuous measurements in the supine posture and over the first 10 s (T1), second 10 s (T2), and 3 min (T3) of 80 degrees head-up tilt (HUT).
Results:
Average % deltaSV measured by IFP at T1 (-11.7 +/- 3.7%) was statistically less than the average % deltaSV measured by TIC at T1 (-21.7 +/- 3.1%), while average % deltaSV measured by IFP at T2 (-16.2 +/- 3.9%) and T3 (-19.1 +/- 3.8%) were not statistically distinguishable from the average % deltaSV measured by TIC at T2 (-21.8 +/- 2.5%), and T3 (-22.6 +/- 2.9%). Correlation coefficients (r2) between IFP and TIC were 0.117 (T1), 0.387 (T2), and 0.718 (T3).
Conclusion:
IFP provides beat-to-beat (real-time) assessment of % deltaSV after 20 s of transition to an orthostatic challenge that is comparable to TIC. IFP technology flown during space missions can be used to assess physiological status and countermeasure effectiveness for orthostatic problems that may arise in astronauts after spaceflight. While the peripherally measured IFP response is delayed, the ease of implementing this monitor in the field is advantageous.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Cardiac Output and Stroke Volume
In an average resting adult male, the typical cardiac output averages...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Regulation of Stroke Volume
Preload refers to the degree of stretch on the heart before it contracts. It's analogous to the stretching of a rubber band; the more it's stretched, the more forcefully it snaps back. This concept is encapsulated in the Frank-Starling law of the...