Comparison of two demand oxygen delivery devices for administration of oxygen in COPD
Stefan Langenhof1, Joachim Fichter
1Paracelsus Hospital, Am Natruper Holz 69, 49076 Osnabrueck, Germany.
Aim:
Demand oxygen delivery systems (DODSs) were developed to secure the mobility of patients requiring oxygen therapy. The aim of the present study was to compare the efficacy of two currently available DODS with continuous oxygen administration (CONT).
Patients:
Thirteen patients with COPD (mean [+/- SD] FEV1, 28 +/- 5.2% predicted; mean P(O2), 56.4 +/- 8.1 mm Hg [breathing room air]).
Intervention:
Treatment for 30 min with CONT at a flow rate of 2 L/min, with the DODSs Oxytron 3 (Weinmann; Hamburg, Germany) or DeVilbiss EX 3000 (Somerset, PA) in random sequence. Arterialized blood samples were obtained from a hyperaemized ear lobe after 15 and 30 min.
Results:
After 15 min, no significant differences in P(O2) or arterial oxygen saturation (Sa(O2)) were observed. In comparison with CONT (mean P(O2), 70.5 +/- 10.4 mm Hg; mean Sa(O2), 94.8 +/- 2.13%), oxygenation with the Oxytron 3 (mean P(O2), 66.3 +/- 10.3 mm Hg; mean Sa(O2), 93.5 +/- 2.6%) was significantly less after 30 min when measured independently by blood gas analysis and pulse oximetry. The DeVilbiss EX 3000 (mean P(O2), 69.1 +/- 12.0 mm Hg; mean Sa(O2), 94.5 +/- 3.2%) and CONT showed no differences.
Conclusions:
P(O2) did not reach equilibrium after 15 min of treatment with the DODSs. The titration of a patient to a DODS is recommended, since simply accepting the manufacturer's information on oxygen equivalent does not guarantee an adequate supply of oxygen.
More Related Videos
05:46A Novel Inhalation Mask System to Deliver High Concentrations of Nitric Oxide Gas in Spontaneously Breathing Subjects
Published on: May 4, 2021
07:09Assessment of Pulmonary Capillary Blood Volume, Membrane Diffusing Capacity, and Intrapulmonary Arteriovenous Anastomoses During Exercise
Published on: February 20, 2017
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Inhaled Medications
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...