The Schwarzschild effect of the dosimetry film Kodak EDR 2
A Djouguela1, R Kollhoff, A Rubach
1Pius-Hospital Oldenburg, Clinic of Radiotherapy and Oncology, Medical Radiation Physics Working Group and University of Oldenburg, Oldenburg, Germany.
Abstract:
The magnitude of the Schwarzschild effect or failure of the reciprocity law has been experimentally investigated for the dosimetry film EDR 2 from Kodak. When the dose rate applied to achieve a given dose was reduced by a factor of 12, the net optical density was reduced by up to 5%. The clinical importance of this effect is negligible as long as the films are calibrated at a value of the dose rate approximately representative of the dose rates occurring in the target volume, but in target regions of strongly reduced dose rate the Schwarzschild effect should be allowed for by a correction of the net optical density.