Evaluation of respiratory inductive plethysmography in the measurement of breathing pattern and PEEP-induced changes
Study Objective:
To assess the accuracy of the respiratory inductive plethysmography in the measurement of PEEP-induced changes in end-expiratory lung volume during mechanical ventilation and its accuracy and stability in the measurement of ventilation during controlled mechanical ventilation and spontaneous breathing.
Design:
An open comparison between two methods using a criterion standard. Either a pneumotachometer (mechanically ventilated patients) or a spirometer (spontaneously breathing subjects) was used as the reference method.
Setting:
Tertiary care center; a multidisciplinary intensive care unit and a metabolic research unit.
Patients:
Six mechanically ventilated, paralyzed postoperative open heart surgery patients, six spontaneously breathing COPD patients, and eight healthy volunteers.
Interventions:
Stepwise increases and reductions of PEEP from zero to 12 cm H2O during controlled mechanical ventilation; repeated validation of the calibration of the respiratory inductive plethysmography (RIP) in both mechanically ventilated and spontaneously breathing subjects.
Measurements And Results:
The baseline drift of the RIP in vitro was 10 ml/150 min and in a ventilated model it was 20 ml/150 min. In mechanically ventilated patients, the mean error of the calibration after 150 min was within +/- 5 percent. Change in end-expiratory lung volume (EELV) during the stepwise increase of PEEP up to 12 cm H2O was 849 +/- 136 ml with the RIP and 809 +/- 125 ml with the pneumotachometer (PT), and during the stepwise reduction of PEEP it was 845 +/- 124 ml and 922 +/- 122, respectively (not significant [NS]. The mean difference between methods in the measurement of change in EELV was -6.6 +/- 3.5 percent during increasing and 6.6 +/- 6.7 percent during decreasing PEEP (NS). Both in mechanically ventilated and spontaneously breathing subjects, the difference between methods was significant for VT and VT/TI. The difference in VT was -2.2 +/- 0.2 percent during mechanical ventilation, -1.1 +/- 0.5 percent in spontaneously breathing COPD patients, and 2.9 +/- 0.4 percent in healthy volunteers (NS between groups).
Conclusions:
The RIP is sufficiently accurate for the measurement of PEEP-induced changes in EELV during controlled mechanical ventilation. The accuracy of tidal volume measurement is similar during mechanical ventilation and spontaneous breathing. The calibration of the RIP is stable enough for bedside monitoring of changes in lung volumes.
Related Concept Videos
Assessment of Ventilation I: Respiratory Rate
A Ventilation assessment is critical for monitoring a patient's health status. Respiration, one of the most accessible vital signs, provides insights into the function of numerous body systems and can indicate serious health issues, such as brainstem injuries from head trauma.
Critical Guidelines for Assessing Ventilation:
Respiratory Volumes and Capacities I
Respiratory Volumes
Tidal Volume (TV) Tidal volume (TV) is the air inhaled or exhaled in a...
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like asthma or COPD,...
Assessment of Ventilation II: Respiratory Depth and Rhythm
Respiratory depth measures the volume of air inhaled or exhaled during a breath. It can vary from shallow to deep and typically remains consistent when a person is at rest or asleep. Occasionally, individuals will automatically inhale deeply, known as sighing, which inflates the lungs with more air than normal breathing.
To assess respiratory depth, observe the degree of chest excursion or movement:
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration can...