Related Experiment Video
Updated: Aug 15, 2026

An Experimental Protocol for Assessing the Performance of New Ultrasound Probes Based on CMUT Technology in Application to Brain Imaging
Published on: September 24, 2017
A system for simultaneously measuring contact force, ultrasound, and position information for use in force-based
Michael R Burcher1, J Alison Noble, Lianghao Han
1Medical Vision Laboratory, Department of Engineering Science, University of Oxford, Oxford OX1 3PJ, UK. michael.burcher@philips.com
Abstract:
During freehand ultrasound imaging, the sonographer places the ultrasound probe on the patient's skin. This paper describes a system that simultaneously records the position of the probe, the contact force between the probe and skin, and the ultrasound image. The system consists of an ultrasound machine, a probe, a force sensor, an optical localizer, and a host computer. Two new calibration methods are demonstrated: a temporal calibration to determine the time delay between force and position measurements, and a gravitational calibration to remove the effect of gravity on the recorded force. Measurements made with the system showed good agreement with those obtained from a standard materials testing machine. The system's uses include three-dimensional (3-D) ultrasound imaging, force-based deformation correction of ultrasound images, and indentation testing.