Related Experiment Video
Updated: Apr 23, 2026

Deciphering and Imaging Pathogenesis and Cording of Mycobacterium abscessus in Zebrafish Embryos
Published on: September 9, 2015
[Contamination of endoscopes and endoscope washers by atypical mycobacteria]
K Negayama1, S Terada, H Kiuchi
1Department of Clinical Laboratory, Kagawa Medical School.
Abstract:
Contamination of endoscopes and endoscope washers by atypical mycobacteria was studied. Large amounts of atypical mycobacteria were detected with high frequency inside endoscopes and endoscope washers. The species of atypical mycobacteria was Mycobacterium chelonae subsp. abscessus. Antibacterial-effects of glutaraldehyde against isolated atypical mycobacteria were checked. Sufficient antibacterial-effect was not obtained by 2% glutaraldehyde solution for endoscope sterilization. However, after frequent manual washing and brushing of endoscopes, by using 3% glutaraldehyde solution and 70% alcohol, all endoscopic instruments were completely decontaminated. We must pay attention to contamination of endoscopes and endoscope washer at least once a month.
Related Concept Videos
Healthcare Associated Infections I: Iatrogenic, Exogenic and Endogenic
HAIs significantly increase the cost of health care. Extended stays in healthcare institutions, increased disability, increased costs of medications, including specialized antibiotics, and prolonged recovery times add to the patient's expenses and the healthcare institution and funding bodies.
Bacterial Phylum Actinobacteria
Atypical Pneumonia
Clinical Significance of Antibiotic Resistance