Related Experiment Video
Updated: Aug 15, 2026

On-line Analysis of Nitrogen Containing Compounds in Complex Hydrocarbon Matrixes
Published on: August 5, 2016
Organotin speciation in environmental matrices by automated on-line hydride generation-programmed temperature
1Departamento de Química e Bioquímica, Faculdade de Ciências da Universidade de Lisboa, Campo Grande Ed. C8, 1749-016 Lisboa, Portugal.
Abstract:
In the present contribution, a new automated on-line hydride generation methodology was developed for dibutyltin and tributyltin speciation at the trace level, using a programmable temperature-vaporizing inlet followed by capillary gas chromatography coupled to mass spectrometry in the selected ion-monitoring mode acquisition (PTV-GC/MS(SIM)). The methodology involves a sequence defined by two running methods, the first one configured for hydride generation with sodium tetrahydroborate as derivatising agent and the second configured for speciation purposes, using a conventional autosampler and data acquisition controlled by the instrument's software. From the method-development experiments, it had been established that injector configuration has a great effect on the speciation of the actual methodology, particularly, the initial inlet temperature (-20 degrees C; He: 150 ml/min), injection volume (2 microl) and solvent characteristics using the solvent venting mode. Under optimized conditions, a remarkable instrumental performance including very good precision (RSD < 4%), excellent linear dynamic range (up to 50 microg/ml) and limits of detection of 0.12 microg/ml and 9 ng/ml, were obtained for dibutyltin and tributyltin, respectively. The feasibility of the present methodology was validated through assays upon in-house spiked water (2 ng/ml) and a certified reference sediment matrix (Community Bureau of Reference, CRM 462, Nr. 330 dibutyltin: 68+/-12 ng/g; tributyltin: 54+/-15 ng/g on dry mass basis), using liquid-liquid extraction (LLE) and solid-phase extraction (SPE) sample enrichment and multiple injections (2 x 5 microl) for sensitivity enhancement. The methodology evidenced high reproducibility, is easy to work-up, sensitive and showed to be a suitable alternative to replace the currently dedicated analytical systems for organotin speciation in environmental matrices at the trace level.
More Related Videos
04:17Analysis of Organochlorine Pesticides in a Soil Sample by a Modified QuEChERS Approach Using Ammonium Formate
Published on: January 20, 2023
07:57Quantitative Detection of Trace Explosive Vapors by Programmed Temperature Desorption Gas Chromatography-Electron Capture Detector
Published on: July 25, 2014
Related Concept Videos
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Types of Detectors-II
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Voltammetry: Stripping Methods
Anodic Stripping Voltammetry (ASV)
ASV is used to determine metals and metalloids at trace levels. It involves two steps: deposition and stripping. First, a negative potential is applied to the...