Ultrasound assisted extraction in quantifying lutein from chicken liver using high-performance liquid chromatography
Ting Sun1, Zhimin Xu, J Samuel Godber
1Department of Food Science, Louisiana State University Agricultural Center, Baton Rouge, 70803, USA. zxu@agctr.lsu.edu
Abstract:
Four sample preparation methods, (1) solvent (SOL), (2) saponification and solvent (SP), (3) ultrasound assisted solvent (UA), and (4) saponification and ultrasound assisted solvent (SP-UA), were used for quantifying lutein in chicken liver samples by HPLC. The lutein concentrations obtained by using SOL, UA, SP, and SP-UA were significantly different with values from 10.4 microg/g (UA) to undetected (SOL). Efficiency of the four different methods for extracting lutein from high to low were the UA, SP, SP-UA, and SOL method. The measured value of lutein in the liver sample using the UA method was approximately two and three times higher than that obtained from the SP and SP-UA method, respectively. The methods with saponification significantly affected the stabilities of lutein in liver samples. The lutein concentration measured with the solvent only method was either much lower than any of the other extraction methods or undetectable. This indicated that little lutein in those samples was in a form that could be extracted directly by solvent. Compared with the saponification method, the ultrasound assisted solvent method could effectively extract lutein from sample matrix and thus avoid chemical degradation reactions, which would be especially important for complex biological tissue such as liver.
More Related Videos
11:59Isolation of Lipoprotein Particles from Chicken Egg Yolk for the Study of Bacterial Pathogen Fatty Acid Incorporation into Membrane Phospholipids
Published on: May 15, 2019
07:33Fluorescent Leakage Assay to Investigate Membrane Destabilization by Cell-Penetrating Peptide
Published on: December 19, 2020
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Instrumentation