Rate of change of oxygen concentration for a large animal circle anesthetic system
Adrian M Solano1, Robert J Brosnan, Eugene P Steffey
1Department of Surgical and Radiological Sciences, School of Veterinary Medicine, University of California, Davis, CA 95616, USA.
Objective:
To describe the effects of changes in circuit volume and oxygen inflow rate on inspired oxygen concentration for a large animal circle anesthetic system.
Study Population:
A large animal circle anesthetic system, a 10 L/min flowmeter, and 20- and 40-L breathing bags.
Procedure:
Circuit volume was determined by a carbon dioxide dilution technique. Oxygen flow rates of 3, 6, and 10 L/min were delivered to the circuit with the large breathing bag, and a flow rate of 6 L/min was used with the small bag. Gas samples were collected during a 20-minute period. The time constant (tau) and half-time (T1/2) were calculated and compared with measured values.
Results:
Mean +/- SEM volume of the breathing circuit with a 20- and 40-L breathing bag was 32.97 +/- 0.91 L and 49.26 +/- 0.58 L, respectively. The tau from measurements was 11.97, 6.10, and 3.60 minutes at oxygen flow rates of 3, 6, and 10 L/min, respectively, for the large breathing bag and 3.73 minutes at a flow rate of 6 L/min for the small breathing bag. The T1/2 was 8.29, 4.22, and 2.49 minutes at oxygen flow rates of 3, 6, and 10 L/min, respectively, for the large breathing bag and 2.58 minutes for the small breathing bag.
Conclusions And Clinical Relevance:
This study emphasizes that there are delays in the rate of increase in the inspired oxygen concentration that accompany use of conventional large animal circle anesthetic systems and low rates of inflow for fresh oxygen.
More Related Videos
06:57Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
09:36Halogenated Agent Delivery in Porcine Model of Acute Respiratory Distress Syndrome via an Intensive Care Unit Type Device
Published on: September 24, 2020
Related Concept Videos
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils, connected...
Bioreactor Controls-II
Chemical Factors Affecting Respiration Centers
CO2 has a potent influence on respiration and is strictly regulated. Under...