Monte Carlo modeling of the response of NRC's 90Sr/90Y primary beta standard
T Palani Selvam1, P R B Saull, D W O Rogers
1Ottawa-Carleton Institute of Physics, Carleton University, Ottawa, KIS 5B6, Canada.
Abstract:
The BEAMnrc/EGSnrc Monte Carlo code system is employed to develop a model of the National Research Council of Canada primary standard of absorbed dose to tissue in a beta radiation field, comprising an extrapolation chamber and 90Sr/90Y beta source. We benchmark the model against the measured response of the chamber in terms of absorbed dose to air, for three different experimental setups when irradiated by the 90Sr/90Y source. For the first setup, the chamber cavity depth is fixed at 0.2 cm and the source-to-chamber distance varied between 11 and 60 cm. In the other two cases, the source-to-chamber distance is fixed at 30 cm. In one case the response for different chamber depths is studied, while in the other case the chamber depth is fixed at 0.2 cm as different thicknesses of Mylar are added to the front surface of the extrapolation chamber. The agreement as a function of distance between the calculated and measured responses is within 0.37% for a variation in response of a factor of 29. In the case of dose versus chamber depth, the agreement is within 0.4% for the ISO-recommended nominal depths of 0.025-0.25 cm. Agreement between calculated and measured responses is very good (between 0.02% and 0.2%) for added Mylar foils of thicknesses up to 10.8 mg cm(-2). For larger Mylar thicknesses, deviations of 0.6%-1.2% are observed, which are possibly due to the systematic uncertainties associated with the restricted collisional stopping powers of air or Mylar used in the calculations. We conclude that our simulation model represents the extrapolation chamber and 90Sr/90Y source with adequate accuracy to calculate correction factors for accurate realization of dose rate to tissue at a depth of 7 mg cm(-2) in an ICRU tissue phantom, despite the fact that the uncertainties in the physical characteristics of the source leave some uncertainty in certain calculated quantities.
Related Concept Videos
Types of Radioactivity
Alpha (α) decay is the emission of an α particle from the nucleus. For example, polonium-210 undergoes α decay:
Nuclear Transmutation
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Atomic Nuclei: Nuclear Spin State Population Distribution
Nuclear Stability
To hold positively charged protons together in the...
Isotopes and Radioisotopes
An isotope containing more...