Initial clinical results with a new needle screen storage phosphor system in chest radiograms
1Institut für klinische Radiologie, Ludwig-Maximilians-Universität München. markus.koerner@med.uni-muenchen.de
Purpose:
To evaluate image quality and anatomical detail depiction in dose-reduced digital plain chest radiograms using a new needle screen storage phosphor (NIP) in comparison to full dose conventional powder screen storage phosphor (PIP) images.
Materials And Methods:
24 supine chest radiograms were obtained with PIP at standard dose and compared to follow-up studies of the same patients obtained with NIP with dose reduced to 50 % of the PIP dose (all imaging systems: AGFA-Gevaert, Mortsel, Belgium). In both systems identical versions of post-processing software supplied by the manufacturer were used with matched parameters. Six independent readers blinded to both modality and dose evaluated the images for depiction and differentiation of defined anatomical regions (peripheral lung parenchyma, central lung parenchyma, hilum, heart, diaphragm, upper mediastinum, and bone). All NIP images were compared to the corresponding PIP images using a five-point scale (- 2, clearly inferior to + 2, clearly superior). Overall image quality was rated for each PIP and NIP image separately (1, not usable to 5, excellent).
Results:
PIP and dose reduced NIP images were rated equivalent. Mean image noise impression was only slightly higher on NIP images. Mean image quality for NIP showed no significant differences (p > 0.05, Mann-Whitney U test).
Conclusion:
With the use of the new needle structured storage phosphors in chest radiography, dose reduction of up to 50 % is possible without detracting from image quality or detail depiction. Especially in patients with multiple follow-up studies the overall dose can be decreased significantly.
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation I: X-ray and CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...