Related Experiment Video
Updated: Aug 14, 2026

Pipeline for Planning and Execution of Transcranial Ultrasound Neuromodulation Experiments in Humans
Published on: June 28, 2024
Spectral image reconstruction for transcranial ultrasound measurement
1Department of Radiology, Harvard Medical School, Brigham and Women's Hospital, Boston, MA 02115, USA.
Abstract:
An approach aimed at improved ultrasound resolution and signal strength through highly attenuating media is presented. The method delivers a series of multiple-cycle bursts in order to construct a discrete spectral (frequency domain) response in one dimension. Cross-correlation of this ultrasound A-mode response with its transmitted signal results in time-localized peaks that correspond to scattering locations. The approach is particularly relevant to the problem of transcranial ultrasound imaging, as it combines numerous smaller signals into a single signal whose net power may exceed that which could be achieved using a single burst. Tests are performed with human skull fragments and nylon-wire targets embedded in a tissue phantom. Skulls are oriented to produce both lateral and shear modes of transcranial propagation. A total of nine locations distributed over three ex vivo human skull samples are studied. Compared with pulsed and chirped signals, results indicate more localized peaks when using the multi-cycle approach, with more accurate positioning when combined with the transcranial shear mode.