[Comparative study between the non-contact pneumotonometer Canon TX10 and the Goldmann tonometer]
H Sánchez-Tocino1, R Bringas-Calvo, D Iglesias-Cortiñas
1Departamento de Oftalmología, Hospital Río Hortega de Valladolid, España. hsanchez@auna.es
Purpose:
To study and contrast the measurement of intraocular pressure (IOP) with an air non-contact tonometer (Canon TX-10) with that obtained with the Goldmann tonometer.
Methods:
Intraocular pressure was measured in 649 subjects with both the pneumotonometer and the Goldmann tonometer. Statistically significant differences between the results obtained by both tonometers and the influence of variables including age, corneal thickness, and corneal curvature were studied.
Results:
The mean age of the study population was 48.5 years, standard deviation (SD) 7.6 years, with a range of 22-68 years. The mean IOP was 15.4, SD 2.7 mm Hg, (range 8-25) with the Goldmann tonometer and 15.6, SD 2.9 mm Hg (range 8-29) with the pneumotonometer. There was a statistically significant difference between the measurements using the air pneumotonometer and the Goldmann tonometer (p<0.001). The mean of the differences between both tonometers was 0.24, SD 2 mm Hg. Seventy-five percent of measurements were in the range of +/-1, 85% in +/-2, and 95% in the Bland and Altman's interval of 4.42, -3.94. These differences were more significant for higher values of IOP, in thicker corneas and in corneas with greater curvature. A correlation between IOP and age was not found (p=0.247).
Conclusions:
The pneumotonometer is a useful and reliable instrument for measuring the IOP providing there is additional checking in cases with borderline values of IOP, or excessive corneal thickness and/or curvature.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)