3-Tesla intraoperative MR imaging for neurosurgery
Walter A Hall1, Walter Galicich, Thomas Bergman
1Department of Neurosurgery, University of Minnesota Medical School and Hennepin County Medical Center, Menniapolis, MN 55455, USA. hallx003@umn.edu
Journal of Neuro-Oncology
|November 30, 2005
Summary
Intraoperative magnetic resonance (MR) image-guided neurosurgery is safe and effective at 3-Tesla (T). This advanced technique, adapted from 1.5-T systems, allows for complex brain surgeries within a high-field MR environment.
Area of Science:
- Neurosurgery
- Medical Imaging
- Magnetic Resonance Imaging
Background:
- Intraoperative magnetic resonance (MR) image-guided neurosurgery has been utilized since 1994.
- Over 750 interventional procedures have been performed using a 1.5-Tesla (T) intraoperative MR imaging system.
- The safety and efficacy of this technique are well-established for in-hospital MR suites.
Purpose of the Study:
- To adapt intraoperative MR image-guided neurosurgery to a new short-bore 3-T MR suite.
- To design and implement a surgical suite enabling procedures entirely within a 3-T MR environment.
- To evaluate the safety and efficacy of neurosurgery at 3-T.
Main Methods:
- All surgical instrumentation, including electrocautery and ultrasonic aspirators, was made MR-compatible.
- Procedures were performed within a 3-T MR environment with strict precautions for minimally ferromagnetic items.
- Cases included drainage and reservoir placement, brain biopsies, and craniotomies for biopsy and resection.
Main Results:
- A cystic craniopharyngioma was successfully treated with aspiration and reservoir placement.
- All five brain biopsies yielded diagnostic tissue.
- Craniotomy for mass resection revealed radiation necrosis; image interpretation was accurate despite increased metallic artifact at 3-T.
Conclusions:
- MR-guided neurosurgery can be safely and effectively performed at 3-T.
- The 3-T surgical environment is comparable to the 1.5-T environment for neurosurgical procedures.
- No adverse events were associated with any procedures performed at 3-T.