Related Experiment Videos
Software-induced variance in two-dimensional gel electrophoresis image analysis.
Asa M Wheelock1, Alan R Buckpitt
1Department of Molecular Biosciences, School of Veterinary Medicine, University of California Davis, Davis, CA, USA. asa@para-docs.org
Electrophoresis
|November 30, 2005
Summary
Quantitative proteomics analysis using two-dimensional gel electrophoresis (2-DE) software can introduce significant variability. This study found that Phoretix 2D Expression software generally produced less variance than PDQuest, especially when background subtraction was omitted.
Area of Science:
- Proteomics
- Biotechnology
- Analytical Chemistry
Background:
- Experimental variability is a known issue in two-dimensional gel electrophoresis (2-DE).
- Postexperimental quantitative analysis using 2-DE software packages is a less studied source of variability.
- Accurate protein quantification is crucial for reproducible proteomics research.
Purpose of the Study:
- To evaluate and compare the performance of two 2-DE analysis software programs: Phoretix 2D Expression v2004 and PDQuest 7.2.
- To determine the impact of different algorithms and image processing parameters on quantitative variance.
- To identify sources of software-induced variance in 2-DE quantitative proteomics.
Main Methods:
- Tested background subtraction and smoothing algorithms on single 2-DE gel images and authentic replicate gel sets (n=5).
- Evaluated variance induced by slight shifts in image boundaries ('cropping area').
- Quantified protein spots using both Expression and PDQuest software.
Main Results:
- Both software packages introduced variance in protein spot quantification, even with identical gel images.
- PDQuest exhibited approximately twice the variance (CV(mean) = 8%) compared to Expression (CV(mean) = 4%) due to image boundary shifts.
- Omitting background subtraction algorithms resulted in the least software-based variance.
- Software-induced variance contributed significantly to the total observed variance in replicate gels.
Conclusions:
- 2-DE gel analysis software is a significant source of variance in quantitative proteomics.
- The choice of software and specific algorithms, particularly background subtraction, can impact the reliability of results.
- Optimizing software parameters and potentially omitting certain algorithms may reduce variance in 2-DE-based quantitative proteomics.