Energy equivalent pressure and total hemodynamic energy associated with the pressure-flow waveforms of a pediatric

William J Weiss1, Branka Lukic, Akif Undar

  • 1Departments of Surgery and Bioengineering, The Pennsylvania State University, The Milton S. Hershey Medical Center, Hershey, PA 17033, USA. bweiss@psu.edu

ASAIO Journal (American Society for Artificial Internal Organs : 1992)
|December 3, 2005
PubMed

Insights

This study evaluated a pediatric ventricular assist device (VAD), finding that shorter systolic duration significantly increased pulsatility (energy equivalent pressure). Lower flow rates also enhanced pulsatility in the pediatric VAD.

Area of Science:

  • Biomedical Engineering
  • Pediatric Cardiology
  • Hemodynamics

Background:

  • Pediatric ventricular assist devices (VADs) are crucial for managing heart failure in children.
  • Optimizing VAD performance requires understanding factors influencing pulsatility, a key hemodynamic parameter.
  • Energy Equivalent Pressure (EEP) quantifies hemodynamic energy and is a critical metric for pulsatility assessment.

Purpose of the Study:

  • To quantify the impact of flow rate and systolic duration on the pulsatility of a pulsatile pediatric VAD.
  • To determine how these parameters affect hemodynamic energy transfer and energy loss within the device's outflow cannula.
  • To provide data for optimizing the design and clinical use of pediatric VADs.

Main Methods:

  • A pulsatile pediatric VAD with a 12 ml stroke volume was tested using a mock circulatory loop.
  • Systemic arterial pressure was maintained at approximately 90/60 mm Hg (mean 75 mm Hg).
  • Energy Equivalent Pressure (EEP) was measured at proximal and distal cannula ends across four flow rates (0.50-1.25 L/min) and two systolic durations (230 ms, 400 ms).

Main Results:

  • Arterial EEP ranged from 5.58% to 8.41% with rapid ejection (230 ms) and 2.33% to 4.20% with slower ejection (400 ms).
  • Higher pulsatility (EEP) was observed at lower flow rates, particularly with shorter systolic durations.
  • Hemodynamic energy loss in the outlet cannula increased significantly with decreased systolic duration but was less sensitive to mean flow rate.

Conclusions:

  • Systolic duration is a critical determinant of pulsatility in this pediatric VAD, with shorter durations yielding higher pulsatility.
  • Flow rate also influences pulsatility, with lower rates enhancing the effect, especially under rapid ejection conditions.
  • Findings highlight the trade-offs between pulsatility and energy loss, informing VAD management and design for pediatric applications.