Related Experiment Video
Updated: Aug 14, 2026

In Vivo Confocal Microscopy in the Diagnosis and Management of Dry Eye: A Focus on Imaging Protocols and Interpretation
Published on: November 11, 2025
Comparison of noncontact infrared and remote sensor thermometry in normal and dry eye patients
1Guru Gobind Singh International Eye Research and Cure Centre, New Delhi, India. gurbax_s@yahoo.com
Purpose:
To evaluate the role of closed chamber infrared and remote sensor thermometry in normal and dry eye patients.
Methods:
The study was conducted on 51 dry eye cases (102 eyes), 26 men and 25 women aged 19 to 65 years (35.36+/-14.36), and 51 normal (102 eyes) age- and sex-matched control subjects. The criteria for dry eye were Schirmer-1 (<10 m/5 min), FTBUT (<10 sec), nd lissamine green score (>2). The remote sensor and infrared thermometry was done in closed chamber around the eye in closed and open eye positions.
Results:
In normal eyes, closed chamber infrared thermometry recorded temperature 34.77+/-0.37 degrees C in closed eye position and 35.02+/-0.39 degrees C in open eye position as compared to 27.91+/-2.46 degrees C in closed eye position and 28.01+/-2.46 degrees C in open position with remote sensor thermometry. The difference in temperature from closed to open position was 0.25+/-0.90 degrees C in infrared thermometry and 0.10+/-0.00 degrees C with remote sensor thermometry, which was statistically significant (p<0.0000). In dry eye, the infrared thermometry recorded 35.08+/-0.61 degrees C temperature in closed eye position and 35.53+/-0.63 degrees C in open eye position as compared to 27.41+/-2.48 degrees C in open and closed eye position with remote sensor thermometry. The difference in temperature from closed to open eye position was 0.45+/-0.14 degrees C (p<0.0000) with infrared thermometry as compared to no change 0.00+/-0.00 degrees C with remote sensor thermometry (p<0.0000).
Conclusions:
Remote sensor thermometry proved better for diagnosis of dry eye disease as it showed no change in temperature under closed chamber in closed and open position (p=0.0000). Infrared thermometry was better in recording the absolute temperature from any point on the eye.
Related Concept Videos
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's forehead...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...