Simultaneous Electrocardiography Recording and Invasive Blood Pressure Measurement in Rats
Published on: January 31, 2019
Accurate blood pressure recording: is it difficult?
1Department of Medicine, Post Graduate Institute of Medical Education and Research, Chandigarh, India. ashish_ritibhalla@yahoo.com
Background:
Blood pressure (BP) measurement is a routine procedure but errors are frequently committed during BP recording. AIMS AND SETTINGS: The aim of the study was to look at the prevalent practices in the institute regarding BP recording. The study was conducted in the Medicine Department at Government Medical College, Chandigarh, a teaching institute for MBBS students.
Methods:
A prospective, observational study was performed amongst the 80 doctors in a tertiary care hospital. All of them were observed by a single observer during the act of BP recording. The observer was well versed with the guidelines issued by British Hypertension Society (BHS) and the deviations from the standard set of guidelines issued by BHS were noted. The errors were defined as deviations from these guidelines.
Statistical Methods:
The results were recorded as percentage of doctors committing these errors.
Results:
In our study, 90% used mercury type sphygmomanometer. Zero error of the apparatus, hand dominance was not noted by any one. Every one used the standard BP cuff for recording BP. 70% of them did not let the patient rest before recording BP. 80% did not remove the clothing from the arm. None of them recorded BP in both arms. In out patient setting, 80% recorded blood pressure in sitting position and 14% in supine position. In all the patients where BP was recorded in sitting position BP apparatus was below the level of heart and 20% did not have their arm supported. 60% did not use palpatory method for noticing systolic BP and 70% did not raise pressure 30-40 mm Hg above the systolic level before checking the BP by auscultation. 80% lowered the BP at a rate of more than 2 mm/s and 60% rounded off the BP to nearest 5-10 mm Hg. 70% recorded BP only once and 90% of the rest re inflated the cuff without completely deflating and allowing rest before a second reading was obtained.
Conclusion:
The practice of recording BP in our hospital varies from the standard guidelines issued by the BHS.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: