Determination of nonsteroidal antiinflammatory drugs in water samples using liquid chromatography coupled with
Jolanta Debska1, Agata Kot-Wasik, Jacek Namiesnik
1Department of Analytical Chemistry, Chemical Faculty, Gdansk University of Technology, Poland. chemanal@pg.gda.pl
Abstract:
An analytical method for the determination of trace levels of six different nonsteroidal antiinflammatory drugs (NSAIDs) in water samples has been developed and validated. Environmentally relevant pharmaceuticals were chosen according to human consumption in Poland. Final analysis of the target compounds was performed by RP LC-diode-array detection-MS, whereas sample preparation included an SPE step. For this SPE step, a number of packing materials, such as LiChrolut RP-18, calixarene, Strata-X, BAKERBOND Narc-2, BAKERBOND Polar Plus, BAKERBOND styrene divinylbenzene-1, and Discovery DSC-18, were used, and their respective advantages and disadvantages in this study were discussed. The RP-18 phase was found to be the most retentive for all analytes. The detection limits for compounds in surface waters were varied from 0.005 for diflunisal to 0.095 microg/L for ibuprofen. The average recoveries of NSAIDs from the surface water samples ranged from 80 up to 103%. RSD value is relatively low (from 4% for fenoprofen up to 8% for ibuprofen). The performance of the method was tested with several environmental water samples.
More Related Videos
09:09Semi-Quantitative Analysis of Peptidoglycan by Liquid Chromatography Mass Spectrometry and Bioinformatics
Published on: October 13, 2020
08:56Quantitative Analysis of the Cellular Lipidome of Saccharomyces Cerevisiae Using Liquid Chromatography Coupled with Tandem Mass Spectrometry
Published on: March 8, 2020
Related Concept Videos
High-Performance Liquid Chromatography: Types of Detectors
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences