Monte carlo dose calculations in conventional thorax fields for 60Co photons
Asghar Mesbahi1, Mahmoud Allahverdi, Hossein Gheraati
1Medical Physics Department, Medical Faculty, Tabriz University of Medical Sciences, Tabriz, Iran.
Purpose:
Our purpose was to apply the MCNP4C Monte Carlo (MC) code for dose calculations in the thorax region and compare the results with those of measurements and a conventional treatment planning system (TPS).
Materials And Methods:
We modeled a Theratron 780E 60Co unit and benchmarked our modeling with percent depth doses (PDDs), beam profiles, and output factors measured in a water phantom. For PDDs and beam profiles, the differences between measurements and MC calculations were less than 1% and 2%, respectively. We used an anatomic thorax phantom for evaluation of a conventional TPS and our MC calculation results.
Results:
In comparing the results of calculations and measurements for our thorax geometries, the errors of conventional and MC methods were 20% and 2%, respectively. For the anterior mediastinal field and large thorax field the accuracy of the conventional method was acceptable, but for small fields of lateral thorax irradiation, the error of the conventional method was as high as to 20%. In all MC calculations, discrepancy from the measurements was less than 2%.
Conclusion:
Our results showed that the MCNP4C MC code could be used in dose calculations in treatment planning for 60Co photon irradiation. In addition, the application of the MC method for dose calculations in radiotherapy with 60Co photons was recommended.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy (PRRT): 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Biological Effects of Radiation
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...