Unexpected Errors in Gas Chromatographic Analysis of Methane Production by Thermophilic Bacteria
Byoung-Kwan Kim1, Lacy Daniels
1Department of Microbiology, University of Iowa, Iowa City, Iowa 52242.
Abstract:
Unexpected errors in methane measurement by gas chromatography occurred when samples at thermophilic temperatures were analyzed. With a standard curve prepared at room temperature (25 degrees C), stoppered bottles incubated and sampled at 37 to 85 degrees C showed more methane upon analysis than bottles incubated at 25 degrees C: values at 50, 63, and 85 degrees C were 109, 126, and 125%, respectively, of the 25 degrees C value. All variation between 4 and 50 degrees C can be explained by the temperature difference between culture bottle and sampling syringe, and the variation of methane concentration can be predicted by the gas law. Between 50 and 63 degrees C, there was a more dramatic rise than predicted by theory. These variations are important to consider if thermophilic methane production is to be measured accurately. Methods to avoid errors are discussed.
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Types of Detectors-II
Microbes and Methanogenesis