Pulse total-hemoglobinometer provides accurate noninvasive monitoring
Eisei Noiri1, Naoki Kobayashi, Yoshiaki Takamura
1Department of Nephrology & Endocrinology, University Hospital, University of Tokyo, Tokyo, Japan. noiri-tky@umin.ac.jp
Objective:
Rapid noninvasive measurement of total hemoglobin would be extremely useful for various clinical situations. This study determined the clinical accuracy and utility for a pulse total-hemoglobinometer using four wavelengths: 660 nm (reduced hemoglobin), 805 nm (isosbestic point), 940 nm (oxygenated hemoglobin), and 1300 nm (water density).
Design:
Clinical trial.
Setting:
University school of medicine.
Patients:
Patients were 122 individuals (age, 18-82 yrs; 49.4 +/- 16.0 yrs [mean +/- SD]), including 71 healthy volunteers, 24 patients undergoing surgery, and 27 patients undergoing hemodialysis.
Interventions:
The hemoglobinometer probe, which simultaneously indicated peripheral oxygen saturation, pulse rate, and hemoglobin, was placed on the fingertip similarly to a regular pulse oximeter. The hemoglobin values were compared with those obtained by the co-oximeter or the sodium lauryl sulfate-methemoglobin method. Those hemoglobin values were assigned to either the training set or the validation set for statistical evaluation.
Measurements And Main Results:
Multiple regression analysis including the ratio of the pulsatile optical density (phi(ij)) derived from the four wavelengths and other factors demonstrated that the mean value of the normalized pulse wave obtained from the photodiode at 805 nm (DC805) and the ratios of DC940 and DC1300 (DC940/DC1300) were the pivotal factors in the hemoglobinometer's increased accuracy in the clinically useful range. The coefficient of determination between both methods was r2 = .81 (p < .0001) in the training set and r2 = .75 (p < .0001) in the validation set. When the cutoff value of anemia was set at 10 g/dL, and anemia was defined as <10 g/dL, the respective sensitivity and specificity of hemoglobinometer values to detect anemia in intraoperative patients were 84.3% and 84.6% (n = 20).
Conclusions:
The data demonstrated the necessity for consideration of light scattering in red blood cells for pulse-spectrophotometric hemoglobin measurement. This was accomplished with additional factors, such as DC805 and DC940/DC1300. With these improvements, the pulse hemoglobinometer provided noninvasive, clinically acceptable measurement of hemoglobin. The pulse hemoglobinometer is a versatile tool that might be useful for routine health checkups of neonates and young children, intraoperative monitoring of bleeding, and emergency care.
More Related Videos
07:51Standardized Hemorrhagic Shock Induction Guided by Cerebral Oximetry and Extended Hemodynamic Monitoring in Pigs
Published on: May 21, 2019
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Measurement of Blood Pressure
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.