Mechanical testing of human cardiac tissue: some implications for MRI safety
Maria-Benedicta Edwards1, Edward R C Draper, Jeffrey W Hand
1United Kingdom Heart Valve Registry, Department of Cardiothoracic Surgery, Hammersmith Hospital, UK.
Purpose:
The effects of aging on tissue strength and its ability to withstand forces associated with MRI have not been investigated. This study aimed to determine the forces required to cause partial or total detachment of a heart valve prosthesis in patients with age-related degenerative diseases exposed to MRI.
Methods:
Eighteen tissue samples excised during routine heart valve replacement surgery were subjected to a suture pull-out test using a tensile materials testing machine. Five preconditioning cycles were applied before commencing the final destructive test. The test was complete when the sample ruptured and the suture was pulled completely free from the tissue. Results were compared with previously calculated magnetically induced forces at 4.7 T.
Results:
All tissue samples displayed a basic failure pattern. Mean forces required to cause initial yield and total rupture were 4.0 N (+/- 3.3 N) and 4.9 N (+/- 3.6 N), respectively. Significant factors determining initial yield were stenosed calcific tissue (p < .01), calcific degeneration (single pathology) (p < .04) and tissue stiffness (p < .01). Calcific degeneration (p < .03) and tissue stiffness (p < .03) were also significant in determining maximum force required to cause total rupture.
Conclusion:
Specific age-related degenerative cardiac diseases stiffen and strengthen tissue resulting in significant forces being required to pull a suture through valve annulus tissue. These forces are significantly greater than magnetically induced < 4.7 T. Therefore, patients with degenerative valvular diseases are unlikely to be at risk of valve dehiscence during exposure to static magnetic field < or = 4.7 T.
More Related Videos
12:24Noninvasive Assessment of Cardiac Abnormalities in Experimental Autoimmune Myocarditis by Magnetic Resonance Microscopy Imaging in the Mouse
Published on: June 20, 2014
06:29Cardiac Magnetic Resonance for the Evaluation of Suspected Cardiac Thrombus: Conventional and Emerging Techniques
Published on: June 11, 2019
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System V: CT