Fluoroscopically guided implantation of modern cardiac resynchronization devices: radiation burden to the patient and
Kostas Perisinakis1, Nicholas Theocharopoulos, John Damilakis
1Department of Medical Physics, Faculty of Medicine, University of Crete, Iraklion, Greece. perisina@med.uoc.gr
Objectives:
To establish radiation risks for patients undergoing fluoroscopically guided cardiac resynchronization device implantation.
Background:
Cardiac resynchonization therapy (CRT) may be associated with extended fluoroscopic exposure.
Methods:
The fluoroscopy time, dose-area product (DAP), exposure parameters, and percentage contribution of the fluoroscopic projections commonly used were recorded in a series of 14 consecutive patients referred for cardiac resynchronization device implantation and compared to corresponding data obtained from a control group of 20 patients who underwent a conventional rhythm device implantation operation. The DAP to peak skin dose, DAP to effective dose, and DAP to gonadal dose conversion factors were determined for biventricular pacing and conventional rhythm device implantation using a humanoid phantom and thermoluminescence dosimetry.
Results:
The mean total fluoroscopy time and DAP values were 35.2 min and 4,765 cGy cm2, respectively, for biventricular pacing and 8.2 min and 1,106 cGy cm2, respectively, for conventional rhythm device implantation. Patient skin dose from biventricular pacing procedures requiring extended fluoroscopic exposure may exceed threshold dose for the induction of skin effects only if X-ray source-to-skin distance is kept low. The risk values for fatal cancer and severe hereditary disorders, respectively, associated with a typical CRT procedure were 273 per million and 0.2 per million treated patients.
Conclusions:
Radiation risks associated with fluoroscopically guided CRT procedures may be considerable. Present data may be used for the estimation of patient radiation risks from CRT procedures performed in other institutions.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System IV: CMRI
Cardiomyopathy IV: Restrictive Cardiomyopathy