Cardiac dysrhythmias after transcatheter closure of ASD with Amplatzer device
Alpay Celiker1, Süheyla Ozkutlu, Cemşit Karakurt
1Units of Pediatric Cardiology, Department of Pediatrics, Hacettepe University Faculty of Medicine, Ankara, Turkey.
Insights
Transcatheter closure of atrial septal defect (ASD) using the Amplatzer device is safe in children. Post-procedure cardiac dysrhythmias are rare and generally benign, with further long-term follow-up needed.
Area of Science:
- Cardiology
- Pediatric Cardiology
- Interventional Cardiology
Background:
- Transcatheter closure is an alternative to open heart surgery for atrial septal defects (ASD).
- Limited data exists on arrhythmia analysis after transcatheter closure of secundum ASD using the Amplatzer device in pediatric patients.
Purpose of the Study:
- To evaluate cardiac dysrhythmias using 24-hour ambulatory electrocardiographic (ECG) monitoring after transcatheter closure of secundum ASD with the Amplatzer device.
Main Methods:
- Eighty-five pediatric patients with ASD underwent transcatheter closure with the Amplatzer device.
- Sixty-five patients were assessed via 24-hour ambulatory ECG monitoring at a mean follow-up of four months.
- Transient atrioventricular (AV) block and junctional rhythm were noted during procedures.
Main Results:
- Five patients (7.6%) showed rare dysrhythmias on Holter monitoring: supraventricular extrasystoles (2), ventricular premature beats (2), and intermittent sinus arrest (1).
- Transient AV block resolved within two hours in one patient.
- No significant late dysrhythmias were observed during the follow-up period.
Conclusions:
- Transcatheter device closure of secundum ASD with the Amplatzer device is associated with rare and benign dysrhythmias in pediatric patients.
- Long-term follow-up is necessary to fully assess the incidence of late dysrhythmias.
Abstract:
Transcatheter closure of atrial septal defect (ASD) has been used as an alternative to open heart surgery. Although transcatheter closure of ASD with the Amplatzer septal occluder is a safe and feasible method in pediatric patients, there is little published data on arrhythmia analysis following transcatheter device closure of secundum ASD. We evaluated cardiac dysrhythmias with 24-hour ambulatory electrocardiographic (ECG) monitoring after transcatheter closure of ASD with Amplatzer device. A total of 85 consecutive patients with ASD underwent transcatheter closure of secundum ASD with Amplatzer device between October 1998 and December 2003. The study involved 65 of these patients assessed by 24-hour ambulatory ECG monitoring. Seven patients were evaluated a second time by 24-hour monitorization. During the procedure, transient complete atrioventricular (AV) block was seen in two patients. One of them returned to normal sinus rhythm in catheterization lab and the other returned to normal sinus rhythm in two hours. Transient junctional rhythm was observed in another patient during the device placement. Twenty-four hour ambulatory ECG monitoring was performed on all patients after a mean four-month period (1-12 months). Holter recordings demonstrated rare supraventricular extrasystole in two patients, rare ventricular premature beats in two patients, and intermittent sinus arrest with sinus pause lasting <1.5 seconds in one patient, for a total of five patients (7.6%). In conclusion, dysrhythmias after transcatheter device closure of secundum ASD with Amplatzer device are rare and benign. We need further long-term follow-up to evaluate late dysrhythmias after the transcatheter device closure of secundum ASD.