Simple pressure measurement is not reliable in detection of access stenosis in native AV fistulas
M Portová1, J Horení, E Kremenová
1Department of Medicine, Prague--Strahov, Czech Republic.
Abstract:
Dynamic arterial and venous pressures (PA, PV) are used as the simplest tools to assess vascular access quality (VAQ). An increased PV over three consecutive dialyses is believed to indicate a stenosis, a rule devised for synthetic grafts (AVG) but not adequately validated for AV fistulas (AVF). In this study dynamic PV and static intra-access pressure (calculated by means of the simplified formula PIA=(PA+PV)/2) changes were evaluated in 46 accesses in which balloon angioplasty had to eventually be performed. The whole group consisted of 30 forearm AVF, 5 upper arm AVF and 11 AVG. Pressures were compared in each patient at a time of satisfactory access flow (QVA) and immediately before the angioplasty and pressure difference over that period (deltaPV, deltaPIA) evaluated. Despite a significant drop in QVA over the follow-up interval in both AVF and AVG, the mean deltaPV and deltaPIA in AVF were only several mm Hg and the chosen threshold limit of 20 mmHg was exceeded in approximately 10% of patients only. The results in the AVG group were, however, very different: The mean deltaPV and deltaPIA were close to 20 mmHg and almost 60% of patients in the AVG group exceeded this limit. Evaluation of PIA did not improve stenosis detection in either group. It is concluded that PV and/or PIA monitoring may be useful to detect a stenosis in AVG but not in AVF.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure
Pre-Procedural Guidelines for Assessing Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...