Related Experiment Video
Updated: Aug 14, 2026

Chemiluminescence-based Assays for Detection of Nitric Oxide and its Derivatives from Autoxidation and Nitrosated Compounds
Published on: February 16, 2022
Evaluation of the chemiluminescent nitrogen detector for solubility determinations to support drug discovery
Shobha N Bhattachar1, James A Wesley, Christopher Seadeek
1Pfizer Global Research and Development, Ann Arbor Laboratories, 2800 Plymouth Road, Ann Arbor, MI 48105, USA. shobha.bhattachar@pfizer.com
Abstract:
Solubility measurements using chemiluminescent nitrogen detection (CLND) has advantages of reduced compound requirement and increased throughput compared to UV-spectrophotometric and HPLC-based measurements. CLND with direct flow injection was evaluated for the measurement of thermodynamic solubility to support drug discovery. The limit of quantitation (LOQ), accuracy, and day-to-day reproducibility of the detector were measured. Measurements made on CLND were compared to those obtained from UV spectrophotometry and HPLC. Based on the results obtained, it was concluded that the CLND performs satisfactorily for discovery stage thermodynamic solubility measurements.
More Related Videos
08:25Measuring Nitrite and Nitrate, Metabolites in the Nitric Oxide Pathway, in Biological Materials using the Chemiluminescence Method
Published on: December 25, 2016
07:38A General Method for Detecting Nitrosamide Formation in the In Vitro Metabolism of Nitrosamines by Cytochrome P450s
Published on: September 25, 2017
Related Concept Videos
Photoluminescence: Applications
High-Performance Liquid Chromatography: Types of Detectors
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...