A wide-aperture dynamically focusing sagittal monochromator for X-ray spectroscopy and diffraction
R L Bilsborrow1, P A Atkinson, N Bliss
1CCLRC Daresbury Laboratory, Daresbury, Warrington WA4 4AD, UK. r.l.bilsborrow@dl.ac.uk
Abstract:
A scanning dynamically focusing sagittal X-ray monochromator accepting 7 mrad of the fan from a 6 T wiggler is in routine use on beamline 16.5 (ultra-dilute spectroscopy) of the SRS at CCLRC Daresbury Laboratory, UK. The energy range covered is 7-27 keV, with a horizontal spot size of <1.1 mm FWHM. Measured monochromatic flux from a Si 220 crystal pair is 1 x 10(11) photons s(-1) (100 mA)(-1) at 9 keV. This level of flux, usually associated with an insertion device on a third-generation source, permits collection of EXAFS data on concentrations at or below 10 ppm.
Related Concept Videos
X-ray Crystallography
Diffraction
Diffraction is the change in the direction of travel experienced by an electromagnetic wave when it encounters a physical barrier whose dimensions are comparable to those of the wavelength of the light. X-rays are electromagnetic radiation with wavelengths about as long as the distance between neighboring...
X-ray Diffraction of Biological Samples
According to Bragg's law, when X-rays strike the sample positioned on a stage, the rays are scattered by the electron clouds around the sample atoms. The X-ray diffraction or scattering is caused by constructive interference of the X-ray waves that reflect off the internal crystal...
Atomic Absorption Spectroscopy: Instrumentation
The atomizer used in AAS can be either a flame atomizer or an...
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used.
Determination of Crystal Structures