Tympanic membrane healing process and biocompatibility of an innovative absorbable ventilation tube
Riccardo D' Eredità1, Roger R Marsh
1Division of Pediatric Otolaryngology, Department of Otorhinolaryngology-Head and Neck Surgery, Vicenza Civil Hospital, Vicenca, Italy. riccardo.deredita@ulssvicenza
Objective:
To investigate tympanic membrane healing process and biocompatibility of pressure-equalizing (PE) tubes, made of a novel biodegradable, absorbable material, in an animal model.
Background:
Myringotomy and pressure-equalizing tube insertion is a frequent otologic procedure in children. However, results may vary because of the unpredictable extrusion rate of tubes. Furthermore, tubes that are no longer required need a surgical procedure to be removed. The ideal pressure-equalizing tube should remain in place for the time selected by the otologist, with no need for subsequent removal. This objective could be met with pressure-equalizing tubes made from a self-disintegrating material.
Methods:
Pressure-equalizing tubes, made of poly-bis(ethylalanate)phosphazene, were inserted in 55 ears of 28 Hartley guinea pigs, with survival times of 10, 30, and 60 days after tube insertion. In vivo reactions between the poly-bis(ethylalanate)phosphazene pressure-equalizing tubes and the tympanic membrane were studied. Tympanic membranes, middle ears, and tubes were examined by scanning electron microscopy.
Results:
There was neither infection nor inflammatory reaction to the tube in any animal. The healing process of tympanic membranes revealed neither residual perforation nor inward spread of skin epithelium into the middle ear cavity. At 30 days, 53% of the tubes had disintegrated. At 60 days, tubes were functioning in 25% of ears.
Conclusion:
These new poly-bis(ethylalanate)phosphazene pressure-equalizing tubes are promising. The healing process of tympanic membranes is excellent, with no complications. The tube disintegration rate can be controlled by varying the formulation of the polymer, obtaining tubes with predictable resorption rates, to adapt treatment to the needs of each single patient.
Related Concept Videos
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...