Structural inhomogeneity in mammography quality control phantoms detected by refraction-enhanced synchrotron
Keiko Imamura1, Norishige Ehara, Yoichi Inada
1Departments of Radiology, St. Marianna University School of Medicine, 2-16-1 Sugao, Miyamae-ku, Kawasaki 216-8511, Japan. keiko401@marianna-u.ac.jp.
Abstract:
Synchrotron radiation imaging with the refraction-enhancement mode visualized structural inhomogeneities in phantoms used for image quality control of mammography. Eight phantoms were examined, all of which were manufactured in the United States and approved by the American College of Radiology as dedicated phantoms. In addition to fiber- and mass-mimicking test objects, each phantom has 5 groups of calcification specks of various sizes. Synchrotron radiation (SR) imaging was performed at Spring-8, a synchrotron radiation facility in Japan. Images were obtained with monochromatic 20-keV x-ray beams, a radiation field of 15 mm X 26 mm at a sample plane, a CCD camera with a resolution of 6 micrometers as a detector, and a sample-to-detector distance of 10 to 11 m. Two hundred and forty specks were evaluated in total in the SR images, and the surrounding area of each speck was also included. Evaluation of the images showed that 14 crack-like structures were depicted near specks, and there were 62 specks with attached void(s) or air bubble(s). Refraction-enhanced SR imaging sensitively detected structural inhomogeneities and abnormalities in phantoms which were implicitly agreed to have a homogeneous matrix and test objects without foreign substances. A possible manufacturing-dependent quality issue was identified. The effect of inhomogeneities detected by SR imaging on visual scoring of specks could not be identified in the tested phantoms; this should be assessed on images of other phantoms in a future study.
Related Concept Videos
X-ray Imaging
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Electron Microscope Tomography and Single-particle Reconstruction
Electron Tomography
Electron tomography can be performed either in TEM or STEM (scanning transmission...