Comparison of effective dose to children and adults from dual X-ray absorptiometry examinations
Glen M Blake1, Marium Naeem, Maria Boutros
1Department of Nuclear Medicine, Guy's, King's and St. Thomas' School of Medicine, Guy's Campus, St. Thomas Street, London SE1 9RT, UK.
Abstract:
Dual X-ray absorptiometry (DXA) is increasingly used to measure bone density in children. If the system software does not include pediatric scan modes, then child examinations must be performed using adult scan modes that give a higher radiation dose to children than adults. This report describes a study to compare the effective dose to children and adults from DXA scans performed on the Hologic Discovery and QDR4500 models. Depth dose measurements were made using thermoluminescent dosimeters in a Rando phantom and were mapped onto the Cristy mathematical phantoms representing a 5-, 10- and 15-year-old child and an adult, and effective dose (ED) was calculated using the ICRP Publication-60 tissue weighting factors. The ED for spine (hip) examinations performed with the Express mode using the default adult scan lengths were 16.1 (9.8), 11.1 (6.7), 5.6 (3.9) and 4.4 (3.1) microSv for a 5-, 10- and 15-year-old child and adult respectively. However, if care is taken to adjust scan lengths appropriately, the child doses were reduced to 9.1 (7.4), 7.1 (5.9) and 5.0 (3.7) microSv. ED figures for the Fast and Array modes were larger by factors of 1.5 and 3 respectively. EDs for whole body scans for a 5-, 10- and 15-year-old child and adult performed on the A-model (W-model) were 5.2 (10.5), 4.8 (9.6), 4.2 (8.4) and 4.2 (8.4) microSv. Using the infant whole body mode (only available on the A-model), they were 7.5 microSv for a 1-year-old and 8.9 microSv for a neonate. Although doses from child DXA examinations are low, it is still important to keep them as small as possible. DXA operators using Discovery systems can do this by using the Express scan mode, by setting appropriate values of the scan length before scan acquisition and by avoiding mistakes that lead to scans having to be unnecessarily repeated.
Related Concept Videos
Drug Dosing: Infants and Children
X-ray Imaging
Biological Effects of Radiation
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation I: X-ray and CT
Pharmacokinetics in Pediatric Patients: Overview and Drug Absorption