A prospective comparison of 3 new-generation pulse oximetry devices during ambulation after open heart surgery
Nancy M Richards1, Karen K Giuliano, Philip G Jones
1Mid America Heart Institute, Saint Luke's Hospital, 4401 Wornall Road, Kansas City, MO 64111, USA. nrichards@saint-lukes.org
Objective:
To assess the clinical performance of 3 new-generation pulse-oximetry signal-processing software systems (Philips FAST, Masimo SET, and Nellcor N-3000) during ambulation after open-heart surgery.
Design:
Prospective, convenience sample.
Setting:
Cardiac surgical progressive care unit in a 629-bed, not-for-profit, tertiary-care teaching hospital.
Patients:
Status post-cardiac-surgery patients (n = 36) during their first postoperative ambulation.
Interventions:
None.
Protocol:
Randomization was used for digit and hand selection, and all 3 devices were used continuously during ambulation. Data on dropouts, false alarms, and correlation with heart rate were recorded. We continuously measured arterial oxygen saturation via pulse oximetry during ambulation with all 3 devices.
Results:
Pairwise comparisons indicated significant differences among the 3 devices for data dropout and false alarm. In repeated-measures analysis, the Nellcor N-3000 had the greatest likelihood of data dropout (odds ratio of 31.9 to Masimo and 5.6 to Philips, at the 95% confidence interval). However, the converse was true for false alarms; the Masimo had the most false alarms, with an odds ratio of 17.9 to Nellcor and 2.3 to Philips, at the 95% confidence interval. There were also significantly more dropouts with all 3 devices when readings were taken on a hand on an arm from which a radial graft had been taken (p = 0.004). For heart-rate correlation, the mean absolute difference among the 3 devices was similar: Philips = 4.3 beats/min, Masimo = 5.1 beats/min, and Nellcor = 3.0 beats/min.
Conclusions:
There are significant differences among the 3 devices with regard to dropout and false alarms. High numbers of dropouts are problematic because no pulse-oximetry patient information is available during dropout. However, false alarms are even more problematic, because they desensitize clinicians to alarms and call into question the accuracy of displayed data. While our data highlight the statistical differences between the studied oximeters, the clinical implications of these differences warrant further study.
More Related Videos
06:57Use of an Integrated Low-Flow Anesthetic Vaporizer, Ventilator, and Physiological Monitoring System for Rodents
Published on: July 9, 2020
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
Related Concept Videos
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac muscle...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...