Rapid micro-scale assay for homocysteine by liquid chromatography-tandem mass spectrometry
Joachim Kuhn1, Christian Götting, Knut Kleesiek
1Institut für Laboratoriums- und Transfusionsmedizin, Herz- und Diabeteszentrum Nordrhein-Westfalen, Universitätsklinik der Ruhr-Universität Bochum, Bad Oeynhausen, Germany. jkuhn@hdz-nrw.de
Objectives:
Increased total homocysteine (tHcy) level is an independent risk factor for atherosclerosis and cardiovascular disease. Here, we describe a rapid tHcy micro-scale assay.
Methods:
We developed an easy sample preparation and rapid liquid chromatography-tandem mass spectrometry (LC-MS/MS) method for the measurement of tHcy in human serum and plasma, respectively.
Results:
The tHcy assay was linear up to 61.6 micromol/L, the detection limit was 1.0 micromol/L, the lower limit of quantification was 1.8 micromol/L, the imprecision in the range of 9.30-25.9 micromol/L was less than 5.9% and the analytical recovery was 94.7 +/- 6.3%.
Conclusions:
Our novel tHcy LC-MS/MS assay is a quick, precise and robust method for tHcy determination in routine diagnostics.
More Related Videos
05:52Histone Modification Screening using Liquid Chromatography, Trapped Ion Mobility Spectrometry, and Time-Of-Flight Mass Spectrometry
Published on: January 12, 2024
14:42Liquid Chromatography Coupled to Refractive Index or Mass Spectrometric Detection for Metabolite Profiling in Lysate-based Cell-free Systems
Published on: September 23, 2021
Related Concept Videos
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
High-Performance Liquid Chromatography: Types of Detectors