Optimization of instrumental parameters of a near-field thermal-lens detector for capillary electrophoresis
Mikhail A Proskurnin1, Svetlana N Bendrysheva, Nataliya Ragozina
1M.V. Lomonosov Moscow State University, Chemistry Department, Vorob'evy Hills d. 1 str. 3 119992 GSP-3 Moscow, Russia. Michael@analyt.chem.msu.ru
Abstract:
The optical scheme of a near-field dual-beam mode-mismatched thermal-lens detector for capillary electrophoresis with a crossed-beam configuration employing a multimode HeCd laser (325 nm) as an excitation source was optimized. It is shown that a multimode laser can be successfully used as an excitation source in thermal lensing with minimal deviations in thermal responses from Gaussian excitation sources. An equation for diffraction thermal-lens theory for near-field measurements is deduced, and the experimental results agree with the deduced equation. The temperature rise in the capillary was estimated, and the exponential decrease of the signal with time for static conditions and low flow velocities was explained. The optimum configuration of the detector from the viewpoint of the maximum sensitivity and beam sizes was found. The detector provides a significant improvement in the detection limits for model compounds absorbing at 325 nm (nitrophenols) compared to the results obtained with a commercial absorbance detector operating at the same wavelength.
Related Concept Videos
Capillary Electrophoresis: Instrumentation
Electrophoresis: Overview
There...
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Optimizing Chromatographic Separations
Band broadening refers to spreading solute bands as they travel through the column. This broadening can impact resolution. Plate height (H) represents the length required for one theoretical plate. A lower plate height corresponds to...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...