Related Experiment Video
Updated: Aug 13, 2026

Ferromagnetic Bare Metal Stent for Endothelial Cell Capture and Retention
Published on: September 18, 2015
Conformation of adjacent self-expanding stents: a cross-sectional in vitro study
Marion Hughes1, Andrew R Forauer, Mats Lindh
1Section of Vascular & Interventional Radiology, University of Michigan Medical Center, Ann Arbor, Michigan, USA.
Abstract:
We examined the proximal conformation of three commonly used self-expanding stents when the stents were deployed adjacent to one another in a tubular model, simulating a "kissing" stent technique. The stent pairs were evaluated by computed tomogrphy to determine the cross-sectional area excluded by the stents within the model. The mean areas associated with each stent pair were compared and significance evaluated by a t-test. A statistically significant difference was found when the area excluded by adjacent Wallstents was compared with both the Luminexx and SMART stents (p < 0.001 and p < 0.002, respectively). The difference in the area excluded and differences in conformation might play a role in the lower patencies that have been observed in "kissing" stent series.