Background:
Internal fixation of fractures using plates and screws is a common method of treatment. Occasionally the internal fixation fails prior to fracture healing. This often requires revision surgery. Determining the force that internal fixation needs to withstand postoperatively would enable this force to be applied intraoperatively as a test to predict successful fixation. The purpose of the present paper was to determine the minimum stripping torque needed to predict successful internal fixation strength.
Methods:
The pull-out strength and stripping torque relationships of 4.5-mm cortical bone screws in Sawbones polyurethane foam were determined. Screw forces were directly measured using an LCM load cell washer on a model intertrochanteric neck of femur fracture fixed with 135 degrees 4-hole pin and plate loaded to single leg stance conditions. Additionally a 135 degrees 4-hole pin and plate was mounted on foam blocks and loaded until failure of the shaft screws from the foam occurred. Predicted stripping torque/yield load was determined.
Results:
Pull-out strength and stripping torque of 4.5-mm cortical bone screws in polyurethane foam have a high degree of linear correlation R(2) = 0.95. Direct measurement of shaft screw forces at single leg stance conditions were 585-686 N. This correlated with a stripping torque of 0.9 Nm. Load to yield testing at single leg stance conditions corresponded to a stripping torque of 1.8 Nm.
Conclusion:
Withstanding 0.9-1.8 Nm of torque during insertion of the femoral shaft screws of a 135 degrees 4-hole pin and plate predicts that the construct will successfully withstand single leg stance.
More Related Videos
11:21Development of a Novel Internal Fixation Model for Rat Radial Fractures: Fracture Healing Assessment and Dorsal Root Ganglion Isolation
Published on: March 13, 2026
07:35Creating Rigidly Stabilized Fractures for Assessing Intramembranous Ossification, Distraction Osteogenesis, or Healing of Critical Sized Defects
Published on: April 11, 2012
Related Concept Videos
Residual Stresses in Circular Shafts
Circular Shafts - Elastoplastic Materials
As torque on the...
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...
Net Torque Calculations
Screw: Problem Solving
To evaluate the...
Stress Concentrations in Circular Shafts