Is stress-only imaging practical when a 1-day stress-rest 99mTc-tetrofosmin protocol is used?
Andrew M Cheetham1, Victoria Naylor, Jean McGhie
1Department of Nuclear Medicine, Harefield Hospital, UK.
Objectives:
To investigate whether a significant number of normal 1-day stress-rest 99mTc-tetrofosmin myocardial perfusion scintigraphy (MPS) studies can be identified from the low-dose stress acquisition alone, and whether technical staff can find such studies reliably.
Methods:
The supervising consultant and four technologists independently graded the stress acquisitions from 200 consecutive MPS studies using a five-point scale. Studies were classified as normal or abnormal according to the final clinical report based on the completed stress-rest protocol.
Results:
Between 31 and 62 studies (16-31%) were classified as definitely normal from the stress acquisition alone, of which 0-4 (0-9%) proved abnormal on the final report. Of stress studies graded definitely normal by each technologist, the consultant disagreed in 13-34% of cases. Of 78 stress studies graded definitely normal by at least one technologist, 6% turned out to be abnormal and the consultant disagreed in 33%. When there was agreement between at least two technologists (57 studies), the rates fell to 4% and 21% respectively.
Conclusions:
Technologists could make the decision to avoid a resting study in up to 30% of patients following a normal low-dose stress acquisition. The supervising nuclear cardiologist would disagree in perhaps one in five cases, even when there was consensus between two or more technologists. These patients would suffer minor inconvenience by being recalled for a rest acquisition on a second day, but there would be important savings in time and radiation exposure for the majority.
More Related Videos
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body being...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies III: Computed Tomography