High-resolution Functional Magnetic Resonance Imaging Methods for Human Midbrain
Published on: May 10, 2012
The cost of parallel imaging in functional MRI of the human brain
Henry Lütcke1, Klaus-Dietmar Merboldt, Jens Frahm
1Biomedizinische NMR Forschungs GmbH am Max-Planck-Institut für Biophysikalische Chemie, 37070 Göttingen, Germany. hluetck@gwdg.de
Abstract:
While the advantages of parallel acquisition techniques for echo-planar imaging (EPI) are well documented for studies affected by magnetic field inhomogeneities, this work focuses on the costs in functional MRI of brain regions without artifacts due to susceptibility effects. For a visual stimulation paradigm and relative to conventional EPI (2.9 T; TR/TE=2000/36 ms), the use of parallel acquisition at a reduction factor of 2 decreased the mean number of activated voxels by 21% at 2 x 2 x 2-mm(3) resolution (n=6) and by 15% at 3 x 3 x 3-mm(3) resolution (n=6). The loss of sensitivity reflects both a decreased signal-to-noise ratio of the native images due to a lower number of contributing gradient echoes and a decreased BOLD MRI sensitivity due to the coverage of a smaller range of TEs.