Comparative in vitro wear testing of PEEK and UHMWPE capped metacarpophalangeal prostheses
T J Joyce1, C Rieker, A Unsworth
1School of Mechanical and Systems Engineering, University of Newcastle upon Tyre, NE1 7RU, UK. t.j.joyce@ncl.ac.uk
Abstract:
Six metacarpophalangeal prostheses were each wear tested to five million cycles. Each prosthesis consisted of a metacarpal component with an approximately hemispherical shell on a titanium body, articulating against a titanium phalangeal component. Four prostheses had a shell made from ultra-high molecular weight polyethylene (UHMWPE) and two had a shell made from polyether ether ketone (PEEK). The tests were undertaken using a finger wear simulator. Despite pre-soaking and the use of control components, lubricant uptake by the metacarpal components was significant. Gravimetrically, the UHMWPE test components showed a greater weight gain than the UHMWPE control components. Therefore there was no apparent wear of any of the UHMWPE test metacarpal components. The original concentric machining marks of the UHMWPE components could still be seen after five million cycles of testing. For the metacarpal components with PEEK shells, gravimetric wear could be measured. Gravimetrically, all of the titanium phalangeal components showed little or no wear. Light scratches in the direction of sliding appeared on the articulating faces of all metacarpal and phalangeal test components, indicating slight abrasive wear.
More Related Videos
08:47Mimicking and Measuring Occlusal Erosive Tooth Wear with the "Rub&Roll" and Non-contact Profilometry
Published on: February 2, 2018
07:42Quasistatic Mechanical Testing for Computer-Aided Design and Manufacturing Occlusal Veneers Cemented to Milled Dentin Analog Material
Published on: December 20, 2024