Related Experiment Video
Updated: Aug 13, 2026

Oral Biofilm Formation on Different Materials for Dental Implants
Published on: June 24, 2018
Retentive force of O-ring attachment to use Immediate Provisional Implant (IPI)-retained overdenture
Mariko Kobayashi1, Chikahiro Ohkubo, Yasunori Suzuki
1Department of Removable Prosthodontics, Tsurumi University School of Dental Medicine, Kanagawa, Japan. kobayashi-mariko@tsurumi-u.ac.jp
Abstract:
This study evaluated the retentive force of the O-ring attachment to an Immediate Provisional Implant (IPI)-retained overdenture. Two sizes of O-rings (#1, #2) were placed on the IPI abutment head. As the controls, soft relining material, silicone lining material, and the PMMA resin were used to connect the IPI abutment head. The retentive forces (n=5, N) obtained at a crosshead speed of 40 mm/min were analyzed by ANOVA/Tukey's HSD test (alpha=0.05). O-ring #1 showed the significantly greatest force among all materials tested (p0.05). Appropriate retention was obtained using the smaller O-ring#1 for the IPI-retained overdenture.