Oxygen conservation during long-distance transport of ventilated patients: assessing the Modified Circle System
1Royal Army, Medical Corps, United Kingdom. tlowes@doctors.org.uk
Introduction:
The Modified Circle System was developed to reduce the amount of oxygen required for long-distance transportation of ventilated patients. Formal testing of the system has never been documented. Various aspects of the system were tested, with the study broken into four parts.
Methods:
Bench testing was performed with a test lung that simulated two different lung compliances. Four ventilators were tested with the system. The four parts to the study were oxygen flow rates, position of oxygen inflow, pressure drop across the system, and specific testing with the LP 10 ventilator.
Results:
The minimum oxygen flows required to maintain forced inspiratory oxygen (FiO(2)) higher than 0.93 over a range of ventilator and lung settings was 1.5 L/min. The oxygen inflow is best connected directly to the CO(2) absorber unit. A significant pressure drop occurs across the system, which increases progressively with higher ventilator pressures. The LP 10 does not allow improved oxygen conservation when recycling gases via its air inlet port.
Conclusion:
The study provides previously unknown, clinically important information regarding the performance of the Modified Circle System.
More Related Videos
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Transport in the Blood
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes:
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is important.
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this principle...