Heat production by 3 implant drill systems after repeated drilling and sterilization
Guillermo E Chacon1, Daniel L Bower, Peter E Larsen
1Department of Oral and Maxillofacial Surgery, College of Dentistry, The Ohio State University, Columbus, OH 43218-2357, USA. chacon.4@osu.edu
Purpose:
The goal was to measure heat generated in bone by 3 implant drill systems after repeated drilling and sterilization.
Materials And Methods:
Temperature was measured with thermocouple technology in vitro using the bovine femoral cortical bone model. Intermittent drilling was accomplished at a constant 2.4-kg load and drill speed of 2,500 rpm. External irrigation at 40 mL/min with normal saline was used. Three implant drill systems-system A (triple twist drills with a relief angle), system B (triple twist drills without a relief angle), and system C (double twist drills with a relief angle)-were evaluated and heat was measured at the final drill in the drilling sequence (4.0 mm or 4.2 mm) at a depth of 15 mm. Thermocouples were placed 0.5 mm from the osteotomy at a depth of 15 mm. Heat measurements were recorded out to 25 uses.
Results:
Results showed temperature increased with multiple uses. System A and C drills had temperature measurements that were below 47 degrees C, even after 25 uses. System B drills had temperatures that exceeded 47 degrees C from the initial use. Light microscopy showed little drill wear even after 25 uses.
Conclusions:
Drill geometry plays a major role in heat production and may explain the increased temperature readings seen in system B. These drills lack relief angles and have the smallest clearance angles of the 3 systems. It also has fewer drills in its drilling sequence compared with systems A and C. This study shows that temperatures increase when drills are used multiple times. Systems A and C had acceptable temperature measurements out to 25 uses. System B drills showed significantly higher heat production with little visual signs of wear.
Related Concept Videos
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Physical Methods for Controlling Microbial Growth: Temperature
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...